Zeiss Axio Observer live cell station manual
Power-up protocol
Only turn on the things you need. The switches
are labeled with numbers.
- If you will be using fluorescence turn on the Xcite lamp,
otherwise leave it off
- Power strip on top shelf
- Power
strip under computer, if you are heating your sample (this
turns on the incubation equipment)
- Microscope (silver push switch on back left)
- QuantEM camera, if you are going to use it
- Start up computer, start MetaMorph, sign in the logbook
The temperature and CO2 controls are via the TFT screen using
the "incubator" tab.
The incubator takes >10 min to come up to temperature and the
system become much more stable over two-hours so it is worth
warming up the system before you start long experiments
The
Microscope
There are 3 stage inserts for holding different samples
- a heated insert for 35 mm glass-bottomed dishes, an insert
for multi-well plates and universal frame for holding slides
and other samples. Carefully insert the one you want to use,
the red dots go together.
Nearly everything on the microscope is motorized
so can be controlled either on the LCD touchscreen or through
MetaMorph.
Objectives
|
Mag |
NA |
Oil? |
DIC? |
|
5x |
0.16 |
NO
|
no
|
|
10x
|
0.30
|
NO
|
Ph1 |
|
20x
|
0.8 |
NO
|
DIC
|
|
40x
|
1.4
|
Yes
|
DIC
|
|
63x
|
1.4
|
Yes
|
DIC
|
|
100x |
1.4 |
Yes |
DIC |
To view your sample down the eyepieces
-
Click "eyepieces" in the MetaMorph task bar
-
Select the illumination you want yo use from the drop down illumination
menu (top left)
-
Click the shutter icon to open/close the shutter (always
keep exposure to a minimum)
-
Coarse/fine focus knobs are on the microscope, xy stage
movement is through the rotary controller, or in MetaMorph
Image acquisition using MetaMorph
Select the camera:
Select the illumination you want to use from the dropdown illumination
menu (eg green, red, DICIII)
Show live to center and focus your sample
Acquisition through the "acquire"
dialog box
This is for simple images of your samples. It can be opened from
the task bar Acquire or from the dropdown menu, Acquire> Acquire
Things
to set in the Acquire window:
- Set
the display to "Autoscale" - the triangles on the
image histogram show the scale min and max.
- The
image gamma - the relationship between measured and displayed
brightness. 1=linear
- Exposure
Time - Type in value (or choose AutoExpose) mouse-over image
to give pixel intensity values (between 0 and 4095)
- Full
chip uses entire CCD camera, sub-arrays produce smaller files
and may be faster
- Binning
- combines CCD pixels to form n x n super pixels - increases
sensitivity but reduces resolution
- Press
Show Live to see the specimen through the camera - it may be
necessary to focus or move the stage slightly
- Clicking
the folder icon allows you to set whether each image starts
in a new window or overwrites the previous image (handy for
optimizing your image, be careful not to lose your data though)
- External
shutter linked to camera should be [Current Shutter]
- Gain
and offset for the camera under the Special tab - Start as
0 and 255 respectively
Press
the Acquire button to capture the image
Saving
images
Simply click save and name your file. Images
will be saved as 12-bit TIFF images with all the original
data. Not all programs will open these images. Photoshop will
open the
files
as 16-bit
images, you will need to adjust the levels ON A COPY of
the file. ImageJ can open them also.
To
save your images as 8-bit files (compatible with all programs)
- Press
Copy to 8-bit button on the task bar
- Select
the desired image from the "Image:" drop-down menu
- Press
Copy and Save by going to File >Save
as (make sure the file name is different otherwise you will
overwrite the original)
Acquisition through the "Multidimensional
acquisition" dialog box
Open
this dialog box for timelapse, multi-wavelength, and
z-stacks for 3D acquisition. You can open the dialog box
from the shortcut on the taskbar.

Main - Select the things you want
to do and then go through
each tab:
Saving
-
Select directory (if you are acquring at a very
high frame rate (eg streaming), use the hard-drive and then move
the data afterwards to your server/memory stick)
-
Choose a base file name (all the images will be named
basename-t1z1 etc)
-
Check the increment base file name box
Timelapse
-
Choose number of time points, duration and interval
(one will always be a consequence of the other two)
Wavelength
- Illumination - choose from the dropdownmenu
- Set exposure time - minimum possible to get
a reasonable image
- Set binning (2 is a good starting point for
many live samples)
- Press the full chip button once before you start to make
sure the entire camera chip is used
- Test the settings using the live and snap buttons at the
bottom of the box (mouse over to label which is which)
Display
- Default is okay for most things, for high frame rates
they should be minimized to not slow down the computer
Summary is an overview of the settings for your experiment
Press the green acquire button to start your multidimensional
acquisition.
The most recent image/stack of each channel will be displayed
as you acquire.
Review multidimensional data
A. Tell metamorph which images you want to build. Press "select
base file" in the main review MDA window which gives you this box
-
- Select directory - point to your folder with the images
- Select data set - choose the individual experiement you want
to look at
- Press View
B. Select the wavelengths, stage positions,
Z-slices and timepoints you want to view.
- Select the stage position (one at a time here) and channels
you want (probably all of them)
- Right clicking in the boxes selects that image. Right click
on the row/column heading selects the entire row/column
- You can choose multi-colour overlay (under display), projection,
or best focus if wanted.
- Use the play and reverse arrows to view the timecouse etc.
- Load Images loads the select images to
a single .stk file which can be saved as one or exported to avi
or quicktime (Stack/Make Movie from the top menu).
Finishing
and shutting down
If there is somebody using the scope after you, clean
the objective(s) you used using lens tissue and cotton
tipped applicators, close metamorph, log-off the computer, fill
in the log book. If the person after you is not going to
use some of the components they can be turned off -
- The fluorescence source (1) can be switched off even if the
rest of the system is on
- Incubator can be turned off as below using the touch pad and
then turning of switch 3 (it's a good idea to then reboot the
microscope for the next person by turning off and on switch
4)
If
there is nobody booked after you within 2 hours, shutdown
the
system in the reverse order of startup, switch 6 to switch
1:
- Using the LCD touch screen, turn off the Heating Inset
P and the CO2 and turn down the XL unit to 20C and open
the incubator doors (This
blows cold air through the system so the heating elements
are not damaged by turning the box off when it is very
hot). After 5 min, when the air coming into the incubator
is
cold,
it
is safe
to
turn
off
the
rest
of the system
- Close Metamorph, shutdown the computer
- Turn off the QuantEM camera, if you used it
- Turn off the incubator powerstrip (under computer), if
you used it
- Turn off the main power strip (on shelf)
- Note the lamp hours in the log book
- Turn off the
fluorescence lamp
|