|
Zeiss Axio Observer manual
Power-up protocol
- Mercury lamp
- Power strip on shelf
- Computer
The Mercury lamp should always be
first-on and last-off. This prevents any electrical surges
caused by ignition damaging other equipment on the same
circuit.
Then log-in with your ASADMIN or local password,
double-click the MetaMorph icon on the desktop and choose
your name from
the drop down menu.
The Microscope
- Binocular eye pieces - adjust the inter-ocular distance
- Wheel on the leftside of the scope controlling the light
path - Either to camera (arrow), eyepice (circle) or 50/50
- Stage - control xy position manually using hanging drive
- Focus knobs - outer part is coarse; inner is fine focus
Objectives
|
Mag
|
NA
|
Oil?
|
DIC/phase
|
|
5x
|
0.16
|
NO
|
Phase1
|
|
10x
|
0.30
|
NO
|
Phase1
|
|
20x
|
0.50
|
NO
|
Phase2
|
|
40x
|
0.75
|
NO
|
Phase2
|
|
63x
|
1.40
|
Yes
|
DIC III
|
|
100x
|
1.40
|
Yes
|
Phase 3
|
Condenser
The condenser is above the stage and is only used for
transmitted light imaging -
H - Brightfield
Phase contrast 1
Phase contrast 2
Phase contrast 3
DIC III
- The DIC analyzer is in filter cube position 1
Transmitted light imaging
- Cube 1 (DIC analyzer) is the best choice for transmitted
images (DIC, phase or brighfield)
- Press the TL button on the scope to turn on the light,
adjust brightness.
Kohler illumination
If you want to accurately observe or capture transmitted light
images, you will need to establish Kohler illumination each
time you change objective. This ensures the illumination
is even and is essential for optimal image quality.
- Focus on your specimen using transmitted light
- Close down the field aperture until it occupies about
2/3 of the field of view.
- Focus the condenser with the knobs underneath the stage
that raise/lower entire condenser. Then center the field
of view using the centering pins also located underneath
the stage. Finally, open the field aperture until it is out
of view.
- Remove eyepiece and close condenser aperture until 2/3
of field is illuminated (or you can adjust this while looking
at your sample until the image has the best contrast)
Fluorescence imaging
The reflector turret contains these filter cubes:
- For transmitted images
- DAPI
- CFP
- GFP
- YFP
- Red
A far-red cube is available.
- Choose the cube you want
- Make sure the TL is off
- Choose "Fluorescence" from the MetaMorph illumination menu
- Click the toggle shutter icon to the right of the menu to
open/close the shutter
- The brightness can be adjusted with the 6-position fluorescence
attentuator
Image acquisition using MetaMorph
When you have located and focused on your specimen, rotate the
wheel to send light to the camera. Remember that the field
of view of the
camera
is a smaller than the field
of view down the eye pieces so make sure your sample is centered.
Open the acquire dialog
box
This can be opened from the task bar Acquire or from the
dropdown menu, Acquire> Acquire
Things to set in the Acquire window:
- Set the display to "Autoscale" - the triangles
on the image histogram show the scale min and max.
- The image gamma - the relationship between measured
and displayed brightness. 1=linear
- Exposure Time - Type in value (or choose AutoExpose)
mouse-over image to give pixel intensity values (between
0 and 4095)
- Full chip uses entire CCD camera, sub-arrays produce
smaller files and may be faster
- Binning - combines CCD pixels to form n x n super pixels
- increases sensitivity but reduces resolution
- Press Show Live to see the specimen through the camera
- it may be necessary to focus or move the stage slightly
- Clicking the folder icon allows you to set whether each
image starts in a new window or overwrites the previous
image (handy for optimizing your image, be careful not
to lose your data though)
- External shutter linked to camera should be [Current
Shutter]
- Gain and offset for the camera under the Special tab
- Start as 0 and 255 respectively
Press the Acquire button to capture the image
Saving images
Simply click save and name your file. Images will
be saved as 12-bit TIFF images with all the original
data. Not all
programs will open these images but you are recommended
to save the original images.
To save your individual images as 8-bit files (compatible
with all programs but some information is lost)
- Press Copy to 8-bit button on the task bar
- Select the desired image from the "Image:" drop-down
menu
- Press Copy and Save by going to File >Save
as (make sure the file name is different otherwise
you will overwrite the original)
If you have more than one colour, you can make an
overlay of the images. Click "overlay images" on
the task bar and follow the dialog box. The result
is 24 bit RGB image
compatible with all programs.
Finishing and shutting down
- Clean any oil objective(s) using lens tissue
and cotton tipped applicators
- Close MetaMorph
- If there is somebody booked after you, logoff the computer
If nobody is scheduled to use the instrument within 1
hour of finishing, please shut it down:
- Computer
- Powerstrip
- Mercury lamp (Remember: First-on, last-off)
- Fill in log book, noting lamp hours
- If you are the last user of the day, please cover the
microscope making sure the cover doesn't touch the hot
mercury lamp housing
|