Zeiss Axio Imager manual
Power-up protocol
- Mercury lamp
- Power strip on shelf
- Computer
The Mercury lamp should always be
first-on and last-off. This prevents any electrical surges
caused by ignition damaging other equipment on the same
circuit.
Then log-in with your "Homer" password, double-click
the MetaMorph icon on the desktop and choose your name from
the drop down menu.
 |
Both the TL and RL shutters on the
microscope should be open (LEDs on). These switches
are on the rightside
at the back of the scope
(The light will now come through when the computer controlled
shutters are activated) |
The Microscope
- Binocular eye pieces - adjust the inter-ocular distance
- Ergotube height adjustment knob - the "periscope"
- Plunger controlling the light path - camera (out) or
eyepieces (in)
- Stage - control xy position manually using hanging drive;
stage clips for holding slide; stage can rotate by loosening
thumbscrew on stage front
- Focus knobs - outer part is coarse; inner is fine focus.
- Z-stepper - this is attached to the left side focus knob.
This controls the z-position allowing auto-focusing and
3D-imaging. The box to the left of the microscope controls
the stepper motor, when it is disengaged manual focus is
possible. When engaged, use ONLY the digipot controller
knob on the box for focusing.
Objectives
|
Mag
|
NA
|
Oil?
|
DIC?
|
|
5x
|
0.15
|
NO
|
no
|
|
10x
|
0.30
|
NO
|
no
|
|
20x
|
0.50
|
NO
|
DIC
|
|
40x
|
0.75
|
NO
|
DIC
|
|
63x
|
1.4
|
Yes
|
DIC
|
|
100x
|
1.4
|
Yes
|
DIC
|
Please change the objectives when the stage is lowered by
carefully turning the knurled ring (rather than pushing the
actual objective).
Condenser
The condenser is beneath the stage and is only used for
transmitted light imaging -
H - Brightfield
D - Dark field, works with low NA objectives (10x,
20x and 40x)
3 - Phase contrast
II - DIC
III - DIC
- The top lens is pushed in using a thumb switch
- The DIC polarizer swings in below the condenser
- The DIC analyzer is in the emission filterwheel or the
manual slider
Transmitted light imaging
- Cube 1 (quad pass dichroic) is the best choice for transmitted
images
- Choose "DIC" or "brightfield" from the drop-down illumination
menu. Click the shutter icon to open/close the shutter.
- If you want to view your sample down the eyepieces as DIC
you will need to have the condenser set appropriately and
have the slider analyzer set to the left (move this back
to the right afterwards or your fluorescence images will
be poor).
Kohler illumination
If you want to accurately observe or capture transmitted light
images, you will need to establish Kohler illumination each
time you change objective. This ensures the illumination
is even and is essential for optimal image quality.
- Focus on your specimen using transmitted light
- Close down the field aperture (dial labeled F on bottom
right of stand) until it occupies about 2/3 of the field
of view.
- Focus the condenser with the knobs underneath the stage
that raise/lower entire condenser. Then center the field
of view using the centering pins also located underneath
the stage. Finally, open the field aperture until it is out
of view.
- Remove eyepiece and close condenser aperture until 2/3
of field is illuminated (or you can adjust this while looking
at your sample until the image has the best contrast)
Fluorescence imaging
The reflector turret contains filter cubes with six positions:
- Quad pass dichroic
- DAPI
- CFP
- GFP
- YFP
- Red
Cubes 2-6 are best for viewing
your sample down the eyepieces. These are also the best choice
for imaging most samples, especially if you have only one fluorophore.
- Choose "CubeXX" from the dropdown illumination menu in
MetaMorph
(where XX is the %power of the excitation)
- Click the toggle shutter icon to the right of the menu
to open/close the shutter
Quad pass dichroic in position 1 is
for taking multi-colour timelapse. It also has a far-red filter
that is not present in any of the individual cubes. From the
the dropdown illumination menu in MetaMorph:
- Choose "DapiXX" for DAPI (where XX is the %power of the excitation)
- Choose "FITCXX" for green fluorophores (eg GFP, FITC, Alexa488)
- Choose "TRITCXX" for red fluorophores (eg Alexa 568, Tritc)
- Choose "CY5XX" for far-red fluorophores (eg Alexa 647 and
CY5)
- Click the toggle shutter icon to the right of the menu to
open/close the shutter
Image acquisition using MetaMorph
When you have located and focused on your specimen manually,
pull out the slider to allow the light to the cameras. Remember
that the field of view of the camera is a smaller than the field
of view down the eye pieces so make sure your sample is centered.
Fluorescence -
Colour camera (eg histology) -
Open the acquire dialog
box
This can be opened from the task bar Acquire or from the
dropdown menu, Acquire> Acquire
Things to set in the Acquire window:
- Set the display to "Autoscale" - the triangles
on the image histogram show the scale min and max.
- The image gamma - the relationship between measured
and displayed brightness. 1=linear
- Exposure Time - Type in value (or choose AutoExpose)
mouse-over image to give pixel intensity values (between
0 and 4095)
- Full chip uses entire CCD camera, sub-arrays produce
smaller files and may be faster
- Binning - combines CCD pixels to form n x n super pixels
- increases sensitivity but reduces resolution
- Press Show Live to see the specimen through the camera
- it may be necessary to focus or move the stage slightly
- Clicking the folder icon allows you to set whether each
image starts in a new window or overwrites the previous
image (handy for optimizing your image, be careful not
to lose your data though)
- External shutter linked to camera should be [Current
Shutter]
- Gain and offset for the camera under the Special tab
- Start as 0 and 255 respectively
Press the Acquire button to capture the image
For colour images
Select "colour" from the illumination
menu, this turns on the transmitted light (no need to toggle
shutter). Make sure the microscope is in kohler illumination.
To take a colour image there is one extra tab
that appears in the acquire dialog box when you click the
colour camera button on the taskbar. . .
These are how the settings in the Acquire and Special tabs should look:
-
Click Autoexpose to find a value that works (you
might need to reduce the lamp brightness if it is too bright)
- Press show live, focus and adjust the xy fine position, then
press measure white balance
- "Acquire" captures a 24bit RGB image. If that looks good you
can save your image and move on to the next slide
If you are struggling with the colour balance:
-
Press the 3200K button on the rightside of the scope (this makes
the bulb very bright and white rather than orange)
- Adjust the ND (neutral density) filters so the intensity
of light at the sample is reasonable
- Capture an image as above
Saving images
Simply click save and name your file. Images will
be saved as 12-bit TIFF images with all the original
data. Not all
programs will open these images but you are recommended
to save the original images.
To save your individual images as 8-bit files (compatible
with all programs but some information is lost)
- Press Copy to 8-bit button on the task bar
- Select the desired image from the "Image:" drop-down
menu
- Press Copy and Save by going to File >Save
as (make sure the file name is different otherwise
you will overwrite the original)
If you have more than one colour, you can make an
overlay of the images. Click "overlay images" on
the task bar and follow the dialog box. The result
is 24 bit RGB image
compatible with all programs.
The colour camera produces 24 bit RGB images directly, which
require no special treatment.
Multidimensional acquisition
Open this dialog box for timelapse and z-stacks for 3D acquisition.
Finishing and shutting down
- Clean any oil objective(s) using lens tissue
and cotton tipped applicators
- Close MetaMorph
- If there is somebody booked after you, logoff the computer
If nobody is scheduled to use the instrument within 1
hour of finishing, please shut it down:
- Computer
- Powerstrip
- Mercury lamp (Remember: First-on, last-off)
- Fill in log book, noting lamp hours
- If you are the last user of the day, please cover the
microscope making sure the cover doesn't touch the hot
mercury lamp housing
|